Skip to the content

FDA Medical Device Recalls (openFDA)

Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc: PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.

Recall number
Z-3186-2018
Recalling firm
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc
Product
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
Status
Terminated
Initiated
2018-04-30
Posted
not on file
Terminated
2019-04-17
Reason
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Root cause
Process control
Action
The firm, Polymedco, initiated a recall by telephone and followed with email on 4/30/18 to inform users to inspect the reagent cartridges for evidence of leakage from the ALP well and to discard cartridges that had evidence of leakage. Polymedco will replace any affected reagent cartridges. If you have any questions, contact Polymedco Technical Support at the 1.800.431.2123 Ext 285.
Quantity
272 KITS in total
Distribution
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Lot, serial or model codes
Lot : A735  Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736  Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736
510(k) numbers
K072189
PMA numbers
not on file
Product code
NBC
Firm FEI
1000119795
Firm city
Cortlandt Manor
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register