Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Sustainability Solutions: REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM

Recall number
Z-3185-2026
Recalling firm
Stryker Sustainability Solutions
Product
REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM
Status
Open, Classified
Initiated
2026-08-19
Posted
2026-09-16
Terminated
not on file
Reason
Reusable access port system may crack or seperate during use.
Root cause
Device Design
Action
On August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Quantity
2682 units
Distribution
US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina , South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Israel and Canada.
Lot, serial or model codes
UDI: (01)00885825020897/ Lots: 16029001 16032878 16038822 16041414 16044000 16048542 16052751 16054119 16062169 16062738 16066292 16074632 16087947 16087952 16090650 16098040 16107246 16111434 16115899 16115909 16120868 16130680 16130701 16135456 16136279 16138395 16140432 16140568 16142544 16142619 16159304 16160563 16172080 16176548 16186738 16186874 16187194 16187248 16187457 16197336 16201786 16205783 16205810 16205823 16206753 16220653 16220918 16221448 16221453
510(k) numbers
["K201511"]
PMA numbers
not on file
Product code
NLM
Firm FEI
3009350690
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register