Skip to the content

FDA Medical Device Recalls (openFDA)

Windstone Medical Packaging, Inc.: AMS Convenience Kit: DSAEK Pack - RX, Part Number AMS12394B

Recall number
Z-3138-2026
Recalling firm
Windstone Medical Packaging, Inc.
Product
AMS Convenience Kit: DSAEK Pack - RX, Part Number AMS12394B
Status
Open, Classified
Initiated
2026-07-29
Posted
2026-09-10
Terminated
not on file
Reason
The convenience kits contain .0.9% Sodium Chloride Injection, USP products in Single Dose freeflex bags, which may contain blue Break-off-Parts (BOPs) that have detached from the infusion port during manipulation.
Root cause
Equipment maintenance
Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 7/29/26 was sent to customers. AMS requests that you undertake the following activities: 1. Immediately check your inventory for the recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for all products affected by this recall. Packs will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE BD announced the recall of 70084 PHARMACY NACL .9% 100ML BAG contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. "At the time the kit is opened for use any BD PHARMACY NACL .9% 100ML BAG should be identified and set aside. "The recalled BD PHARMACY NACL .9% 100ML BAG should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. 3.Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity. 4.If you have further distributed this product, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5.Please complete the Recall Reply Form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand. 6.Indicate on the Recall Reply Form if replacement PHARMACY NACL .9% 100ML BAG are needed. Please complete the attached AMS Recall Reply Form and return it to Aligned Medical Solutions at the Fax number provided on the form. If you do not have the affected product, you must complete and return the AMS Recall Reply Form. Replacement BD Fresenius Kabi Freeflex IV Container contained within the kit will be issued upon request and completion of the attached Recal
Quantity
3,260 kits
Distribution
US Nationwide distribution in the states of CA, GA, WA.
Lot, serial or model codes
UDI-DI:B098AMS12394B0; Lot Numbers: 241907, 241223, 241034, 240904, 240705, 240633, 240269, 238361, 238197, 238146, 238085, 238084, 238041, 237957, 237901, 237843, 237663, 237444, 237443, 236863, 236740, 236250, 236249, 236111, 235602, 235601, 235442, 235026, 235025, 235024, 235023, 234889, 234839, 234458, 233931, 233930, 233762, 233328, 233224, 231945, 231124, 231123, 231122, 231081, 231080, 230970, 230969, 230968, 230697, 230696, 230695, 230694, 230098, 229977, 229825, 229824, 229656, 229655, 229565, 229390, 229238, 229141, 229140, 228793, 228718, 228572, 228263, 227828, 227540, 227112, 226969, 226897, 226841, 226714, 226572, 226382, 225455, 225276, 225275, 223916, 223915, 223914, 223913
510(k) numbers
not on file
PMA numbers
not on file
Product code
KDD
Firm FEI
1000125955
Firm city
Billings
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register