Skip to the content

FDA Medical Device Recalls (openFDA)

Kamiya Biomedical Company, LLC: K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use

Recall number
Z-3138-2017
Recalling firm
Kamiya Biomedical Company, LLC
Product
K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use
Status
Terminated
Initiated
2017-07-19
Posted
not on file
Terminated
2018-11-20
Reason
The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.
Root cause
Under Investigation by firm
Action
Kamiya Biomedical Company sent an Urgent Medical Device Recall/Correction Notice on July 14, 2017 to affected customers, providing them with updated package inserts with corrected A1AT values and/or free replacement kits. Customers were asked to fill in and return the acknowledgement form. Customers with questions were instructed to contact the firm at 206-575-8068.
Quantity
1398 units
Distribution
Worldwide Distribution - US including FL, CA, WA, OH, IA, IL, NJ, NY, NC, MA, TX, CT, MD, VA, Internationally to TN, Czech Republic, Taiwan, Dominican Republic, Switzerland, China, and Jamaica
Lot, serial or model codes
Lot Numbers: N653, N654, D755, D647, F648, F649, G650, J651, L652
510(k) numbers
["K993853"]
PMA numbers
not on file
Product code
JIX
Firm FEI
2084025
Firm city
Tukwila
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register