FDA Medical Device Recalls (openFDA)
Kamiya Biomedical Company, LLC: K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use
- Recall number
- Z-3138-2017
- Recalling firm
- Kamiya Biomedical Company, LLC
- Product
- K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use
- Status
- Terminated
- Initiated
- 2017-07-19
- Posted
- not on file
- Terminated
- 2018-11-20
- Reason
- The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.
- Root cause
- Under Investigation by firm
- Action
- Kamiya Biomedical Company sent an Urgent Medical Device Recall/Correction Notice on July 14, 2017 to affected customers, providing them with updated package inserts with corrected A1AT values and/or free replacement kits. Customers were asked to fill in and return the acknowledgement form. Customers with questions were instructed to contact the firm at 206-575-8068.
- Quantity
- 1398 units
- Distribution
- Worldwide Distribution - US including FL, CA, WA, OH, IA, IL, NJ, NY, NC, MA, TX, CT, MD, VA, Internationally to TN, Czech Republic, Taiwan, Dominican Republic, Switzerland, China, and Jamaica
- Lot, serial or model codes
- Lot Numbers: N653, N654, D755, D647, F648, F649, G650, J651, L652
- 510(k) numbers
- ["K993853"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 2084025
- Firm city
- Tukwila
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28