Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality

Recall number
Z-3114-2017
Recalling firm
Beckman Coulter Inc.
Product
Coulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Status
Terminated
Initiated
2017-06-26
Posted
not on file
Terminated
2023-09-14
Reason
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Root cause
Labeling Change Control
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated June 30, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff. If they forwarded any of the affected product to another laboratory, please provide them a copy of the letter. Customers with questions should contact the Customer Support Center at 800-528-7694.
Quantity
3326 units
Distribution
Worldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, india, Indonesia, Italy, Japan, Iraq, Jordan. Kenya, Kuwait, Libya, Lithuania, Macau, Lebanon, Macedonia, Malawi, Malaysia, Mexico, Mongolia, Morocco, Mozambique, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Paraguay, Philippines, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, UAE, UK, Uruguay, Venezuela, Vietnam, Virgin Islands, Yemen, Zambia, Zimbabwe.
Lot, serial or model codes
Catalog Number(s): HMX CP - 178832, 178833, 178834, A91062, A90994; HmX AutoLoader - 6605522, 6605523, 6605524; All software versions.
510(k) numbers
["K010765"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register