FDA Medical Device Recalls (openFDA)
Ameditech Inc: Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 60503s-3ZC/ PH-10132 60600/ PH-11005
- Recall number
- Z-3051-2020
- Recalling firm
- Ameditech Inc
- Product
- Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 60503s-3ZC/ PH-10132 60600/ PH-11005
- Status
- Terminated
- Initiated
- 2020-08-18
- Posted
- not on file
- Terminated
- 2024-09-30
- Reason
- Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
- Root cause
- Process control
- Action
- A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
- Quantity
- 3600 units
- Distribution
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Lot, serial or model codes
- ALL Lots Within Expiry
- 510(k) numbers
- ["K113046","K113501"]
- PMA numbers
- not on file
- Product code
- DIO
- Firm FEI
- 3003789989
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28