Boston Scientific Corporation: Encore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d) M0067101140

Recall number
Z-3043-2017
Recalling firm
Boston Scientific Corporation
Product
Encore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d) M0067101140
Status
Terminated
Initiated
2017-06-15
Posted
not on file
Terminated
2020-07-22
Reason
Boston Scientific (BSC) has received complaints for cracked trays on the Encore device.
Root cause
Under Investigation by firm
Action
The firm initiated their recall on 06/15/2017 by letter delivered by Fed Ex in the US, on 06/15-29/2017 in the Americas delivered by Courier, email and hand delivery, on 06/16/2017 in Europe by DHL, on 06/16-19/2017 in the Arab Emirates and East Asia by email and hand delivery and on 06/136/2017 in Japan by hand delivery. In the notice, the firm described the problem and requested the following action: 1. Immediately check the inventory and identify whether there are products with UPNs and lot numbers affected by the recall. 2. Inspect the device tray in the area highlighted above. If any units with a cracked tray are identified in the supply, immediately discontinue use and segregate affected units in a secure location. 3. Arrange for the damaged units to be returned to Boston Scientific by completing the attached Reply Verification Tracking Form. 4. If no damage is identified, the device may be used as normal. Please complete and return the attached Reply Verification Tracking Form
Quantity
305,779 units
Distribution
Worldwide
Lot, serial or model codes
Serial Numbers: (a) 19563006, 19587644, 19623142, 19650458, 19704550, 19725489, 19769215, 20054530, 20077271, 20083051, 20127928, 20159518, 20202728, 20206930, 20242335, 20285571, 20309573, 20309574, 20336380, 20394218, 20423644, 20423645, 20456071, 20481170, 20534557  (b) 19640750, 19667072, 19704551, 19755302, 19755307, 19785021, 19808183, 20030825, 20060033, 20087896, 20117909, 20159601, 20171449, 20206929, 20233700, 20256035, 20279769, 20303751, 20336385, 20336386, 20364928, 20364929, 20364930, 20423650, 20472532, 20493563, 20507373, 20534581  (c) 20157078, 19623019, 19755304, 20056496, 20146417, 20264032, 20456079  (d) 19594194, 19717352, 19808180, 20121252, 20223176, 20336387, 20394350, 20534582
510(k) numbers
["K955869"]
PMA numbers
not on file
Product code
MAV
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register