FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories
- Recall number
- Z-3035-2026
- Recalling firm
- Olympus Corporation of the Americas
- Product
- POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories
- Status
- Open, Classified
- Initiated
- 2026-08-18
- Posted
- 2026-08-26
- Terminated
- not on file
- Reason
- Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
- Root cause
- Labeling Change Control
- Action
- Olympus notified consignees on 08/18/2026 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were notified of the issue, hazard involved, and that the affected device could still be used. Consignees were instructed to review the recall notification and current inventory for the affected lot number, discard the outer box for any affected units on hand, and attach the recall notification letter to the inner device packaging. Consignees were also instructed to ensure all affected personnel are made aware, to ensure to check the sterile packaging and device labels before use to ensure you are using the correct length for your procedure, notify customers if products were further distributed and complete and return the provided response form.
- Quantity
- 335 units
- Distribution
- Worldwide - US Nationwide distribution in the states of Kansas and Florida. The countries of Australia, European Union, Japan.
- Lot, serial or model codes
- Model Number: PS-0523CJDA. UDI-DI: 00821925044531. Lot Number: FR490902.
- 510(k) numbers
- ["K212643"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28