Spacelabs Healthcare, Llc: Ultraview Spacelabs Medical Multigas Analyzer, Model 91518, V1.00.14. Spacelabs Healthcare Llc. Monitors gas concentrations and alerts clinical personnel when the concentration of anesthetic agents, oxygen, carbon dioxide, or nitrous oxide falls outside of defined limits. The anesthetic agent being administered is automatically identified.

Recall number
Z-2902-2011
Recalling firm
Spacelabs Healthcare, Llc
Product
Ultraview Spacelabs Medical Multigas Analyzer, Model 91518, V1.00.14. Spacelabs Healthcare Llc. Monitors gas concentrations and alerts clinical personnel when the concentration of anesthetic agents, oxygen, carbon dioxide, or nitrous oxide falls outside of defined limits. The anesthetic agent being administered is automatically identified.
Status
Terminated
Initiated
2011-06-08
Posted
2011-08-01
Terminated
2011-09-21
Reason
Spacelabs received a report of the Model 91518 Multigas Analyzer trending low levels of an anesthetic agent that is not present during a particular case. The error is limited to only the trend table data.
Root cause
Software design
Action
Spacelabs Healthcare, Llc. sent a "Customer Bulletin" dated June 20, 2011 to all affected customers. The customers were notified of the trend problem and are advised to follow the instructions in the Perioperative section and the ADMIT/DISCHARGE section of the users manual. For questions contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support or 1-425-657-7200 x5089.
Quantity
481 units worldwide (333 units internationally and 148 units in the US)
Distribution
Worldwide Distribution: (USA) Nationwide distribution, and the countries of ARGENTINA, BAHRAIN, BOLIVIA, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, ECUADOR, EGYPT, FRANCE, INDIA, MEXICO, NICARAGUA, PARAGUAY, PERU, POLAND, Slovakia, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, CHINA, THAILAND, UNITED ARAB EMIRATES, U.K., URUGUAY, VENEZUELA, and VIETNAM.
Lot, serial or model codes
Model 91518, V1.00.14.  Serial Numbers: 1518-000003, 1518-000008, 1518-000013, 1518-000014, 1518-000016, 1518-000024, 1518-000029, 1518-000036, 1518-000040, 1518-000044, 1518-000226, 1518-000228, 1518-000231, 1518-000233, 1518-000243, 1518-000349, 1518-000354, 1518-000355, 1518-000389, 1518-000399, 1518-000457, 1518-000752, 1518-000819, 1518-000825, 1518-000846, 1518-000948, 1518-000991, 1518-001341, 1518-001393, 1518-001394, 1518-001395, 1518-001522, 1518-001536, 1518-001654, 1518-001655, 1518-001708, 1518-001734, 1518-001736, 1518-001853, 1518-001972, 1518-002065, 1518-002089, 1518-002097, 1518-002114, 1518-002116, 1518-002117, 1518-002118, 1518-002119, 1518-002120, 1518-002121, 1518-002122, 1518-002123, 1518-002124, 1518-002125, 1518-002126, 1518-002127, 1518-002128, 1518-002129, 1518-002130, 1518-002131, 1518-002132, 1518-002133, 1518-002134, 1518-002135, 1518-002136, 1518-002137, 1518-002138, 1518-002139, 1518-002140, 1518-002141, 1518-002142, 1518-002143, 1518-002144, 1518-002145, 1518-002146, 1518-002147, 1518-002148, 1518-002149, 1518-002150, 1518-002151, 1518-002152, 1518-002153, 1518-002154, 1518-002155, 1518-002156, 1518-002157, 1518-002161, 1518-002162, 1518-002163, 1518-002164, 1518-002166, 1518-002167, 1518-002168, 1518-002169, 1518-002170, 1518-002171, 1518-002172, 1518-002173, 1518-002174, 1518-002175, 1518-002176, 1518-002177, 1518-002178, 1518-002179, 1518-002180, 1518-002181, 1518-002182, 1518-002183, 1518-002184, 1518-002185, 1518-002186, 1518-002187, 1518-002188, 1518-002190, 1518-002191, 1518-002192, 1518-002205, 1518-002206, 1518-002207, 1518-002208, 1518-002209, 1518-002210, 1518-002211, 1518-002212, 1518-002213, 1518-002214, 1518-002215, 1518-002216, 1518-002217, 1518-002218, 1518-002219, 1518-002220, 1518-002221, 1518-002222, 1518-002223, 1518-002224, 1518-002225, 1518-002226, 1518-002227, 1518-002228, 1518-002229, 1518-002230, 1518-002231, 1518-002232, 1518-002233, 1518-002234, 1518-002235, 1518-002236, 1518-002237, 1518-002238, 1518-002239, 1518-002240, 1518-002241, 1518-002242, 1518-002243, 1518-002244, 1518-002245, 1518-002246, 1518-002247, 1518-002248, 1518-002249, 1518-002250, 1518-002251, 1518-002252, 1518-002253, 1518-002254, 1518-002255, 1518-002256, 1518-002257, 1518-002258, 1518-002259, 1518-002260, 1518-002261, 1518-002262, 1518-002263, 1518-002264, 1518-002265, 1518-002266, 1518-002267, 1518-002268, 1518-002276, 1518-002277, 1518-002278, 1518-002279, 1518-002280, 1518-002281, 1518-002282, 1518-002283, 1518-002284, 1518-002285, 1518-002286, 1518-002287, 1518-002288, 1518-002289, 1518-002290, 1518-002291, 1518-002292, 1518-002293, 1518-002294, 1518-002295, 1518-002296, 1518-002297, 1518-002298, 1518-002299, 1518-002300, 1518-002301, 1518-002302, 1518-002303, 1518-002304, 1518-002305, 1518-002306, 1518-002307, 1518-002308, 1518-002309, 1518-002310, 1518-002311, 1518-002312, 1518-002313, 1518-002314, 1518-002315, 1518-002316, 1518-002317, 1518-002318, 1518-002319, 1518-002320, 1518-002321, 1518-002322, 1518-002323, 1518-002325, 1518-002326, 1518-002327, 1518-002328, 1518-002329, 1518-002330, 1518-002331, 1518-002332, 1518-002333, 1518-002334, 1518-002335, 1518-002336, 1518-002337, 1518-002338, 1518-002339, 1518-002340, 1518-002341, 1518-002342, 1518-002343, 1518-002344, 1518-002345, 1518-002346, 1518-002347, 1518-002348, 1518-002349, 1518-002350, 1518-002351, 1518-002352, 1518-002353, 1518-002354, 1518-002355, 1518-002356, 1518-002357, 1518-002358, 1518-002359, 1518-002360, 1518-002361, 1518-002362, 1518-002363, 1518-002364, 1518-002365, 1518-002366, 1518-002367, 1518-002368, 1518-002369, 1518-002370, 1518-002371, 1518-002372, 1518-002373, 1518-002374, 1518-002375, 1518-002376, 1518-002377, 1518-002378, 1518-002379, 1518-002380, 1518-002381, 1518-002382, 1518-002383, 1518-002384, 1518-002385, 1518-002386, 1518-002387, 1518-002388, 1518-002389, 1518-002390, 1518-002391, 1518-002392, 1518-002393, 1518-002394, 1518-002395, 1518-002396, 1518-002397, 1518-002398, 1518-002399, 1518-002400, 1518-002401, 1518-002402, 1518-002403, 1518-002404, 1518-002405, 1518-002406, 1518-002407, 1518-002408, 1518-002409, 1518-002410, 1518-002411, 1518-002412, 1518-002413, 1518-002414, 1518-002415, 1518-002416, 1518-002417, 1518-002418, 1518-002419, 1518-002420, 1518-002421, 1518-002422, 1518-002423, 1518-002424, 1518-002425, 1518-002426, 1518-002427, 1518-002428, 1518-002429, 1518-002430, 1518-002431, 1518-002432, 1518-002433, 1518-002434, 1518-002435, 1518-002436, 1518-002437, 1518-002438, 1518-002439, 1518-002440, 1518-002441, 1518-002442, 1518-002443, 1518-002444, 1518-002445, 1518-002446, 1518-002447, 1518-002448, 1518-002451, 1518-002452, 1518-002453, 1518-002454, 1518-002456, 1518-002457, 1518-002458, 1518-002459, 1518-002460, 1518-002461, 1518-002462, 1518-002463, 1518-002464, 1518-002465, 1518-002466, 1518-002467, 1518-002468, 1518-002469, 1518-002470, 1518-002471, 1518-002472, 1518-002476, 1518-002477, 1518-002478, 1518-002479, 1518-002480, 1518-002481, 1518-002482, 1518-002483, 1518-002484, 1518-002485, 1518-002486, 1518-002487, 1518-002488, 1518-002489, 1518-002490, 1518-002491, 1518-002493, 1518-002494, 1518-002495, 1518-002496, 1518-002497, 1518-002498, 1518-002499, 1518-002500, 1518-002501, 1518-002502, 1518-002507, 1518-002508, 1518-002510, 1518-002511, 1518-002513, 1518-002514, 1518-002517, 1518-002518, 1518-002519, 1518-002520, 1518-002521, 1518-002522, 1518-002523, 1518-002524, 1518-002525, 1518-002526, 1518-002527, 1518-002528, 1518-002529, 1518-002530, 1518-002531, 1518-002532, 1518-002533, 1518-002534, 1518-002535, 1518-002536, 1518-002537, 1518-002538, 1518-002539, 1518-002540, 1518-002541, 1518-002542, 1518-002543, 1518-002545, 1518-002549, 1518-002556, 1518-002557, 1518-002558, 1518-002559, 1518-002560, 1518-002561, 1518-002562, 1518-002563, 1518-002564, 1518-002565, 1518-002566, 1518-002567, 1518-002568, 1518-002569, 1518-002570, 1518-002571, 1518-002572, 1518-002573, 1518-002574, 1518-002575, 1518-002576, 1518-002577, 1518-002578, 1518-002579, 1518-002580, 1518-002581, 1518-002582, 1518-002583, 1518-002584, 1518-002585, 1518-002586, 1518-002587, 1518-002588, 1518-002589, 1518-002590, 1518-002591, 1518-002595, 1518-002606, 1518-002607, 1518-002608, 1518-002609, 1518-002610, 1518-002611, 1518-002612, 1518-002613, 1518-002614,  and 1518-005744.
510(k) numbers
["K053599"]
PMA numbers
not on file
Product code
CBR
Firm FEI
3010157426
Firm city
Issaquah
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register