Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Medical Optics Inc. (AMO): TECNIS Multifocal Foldable Acrylic Intraocular Lens, Model No. ZMA00 The product is indicated for primary implantation for the visual correction of aphakia in adult patients with and without presbyopia in whom a cataractous lens has been removed by extracapsular cataract extraction and who desire near, intermediate, and distance vision with increased spectacle independence. The lens is intended to be placed in the capsular bag

Recall number
Z-2863-2016
Recalling firm
Abbott Medical Optics Inc. (AMO)
Product
TECNIS Multifocal Foldable Acrylic Intraocular Lens, Model No. ZMA00 The product is indicated for primary implantation for the visual correction of aphakia in adult patients with and without presbyopia in whom a cataractous lens has been removed by extracapsular cataract extraction and who desire near, intermediate, and distance vision with increased spectacle independence. The lens is intended to be placed in the capsular bag
Status
Terminated
Initiated
2016-08-26
Posted
2016-09-21
Terminated
2016-12-23
Reason
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
Root cause
Process control
Action
Abbott Medical Optics, Inc. sent an Urgent Field Safety Notice to all affected customers to inform them that Abbott Medical Optics (AMO) is recalling 737 IOLs due to detection of an inspection equipment malfunction that resulted in a total of 737 distributed lenses not being fully checked prior to release. The letter informs customers that this malfunction may result in the release of mislabeled IOLs. The letter informs the customers of the actions to be taken. Customers are instructed to complete Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to RegCompliOne@abbott.com. Customers with any questions related to the IOL return process are instructed to contact an AMO Customer Service Representative at 877-266-4543. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543.
Quantity
41 units
Distribution
Worldwide Distribution - US nationwide, PR, and Internationally to Canada, Latin America, Europe, Africa, Middle East, and Asia Pacific.
Lot, serial or model codes
Serial No. 8213991504 8214001504 8214011504 8214021504 8214031504 8214041504 8214051504 8214061504 8214071504 8214101504 8538801603 8538821603 8538831603 8538841603 8538871603 8538891603 8538901603 8538971603 8538981603 8416821507 8416831507 8416841507 8416851507     8416911507 8416921507 8416951507 8416961507 8416971507 8416991507 8417001507 8417011507 8417021507 8417041507 8417071507 8417081507 8417091507 8417101507 8445191507 8505901605 8505911605 8505961605
510(k) numbers
not on file
PMA numbers
P980040
Product code
HQL
Firm FEI
3003843509
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register