Skip to the content

FDA Medical Device Recalls (openFDA)

Dental Equipment LLC: LuxStar Dental Light have the following Agency model or Catalog Model numbers as ML1000, OL1000, CL1000, WL1000, CB1000, TL1000, and UL1000. The product is labeled in part: " Marus Dental***Mfg By: Dental Equipment LLC.***705 S. Springbrook Rd Building C Newberg, Oregon, 97132 USA***Dental Light***MN ML1000***". Product Usage: The intended use of the dental light is used for illuminating the oral cavity during the performance of dental procedures.

Recall number
Z-2834-2011
Recalling firm
Dental Equipment LLC
Product
LuxStar Dental Light have the following Agency model or Catalog Model numbers as ML1000, OL1000, CL1000, WL1000, CB1000, TL1000, and UL1000. The product is labeled in part: " Marus Dental***Mfg By: Dental Equipment LLC.***705 S. Springbrook Rd Building C Newberg, Oregon, 97132 USA***Dental Light***MN ML1000***". Product Usage: The intended use of the dental light is used for illuminating the oral cavity during the performance of dental procedures.
Status
Terminated
Initiated
2011-05-27
Posted
2011-07-15
Terminated
2012-10-24
Reason
Lens heat shield (birdcage) of the Luxstar Dental Light could fall and injure a patient.
Root cause
Device Design
Action
Dental Equipment, LLC sent an "URGENT MEDICAL DEVICE RECALL" notification letter via USPS 1st class to US consignees and via Federal Express or Fax to foreign consignees on June 24, 2011. The letter informed customers about the product, problem and tether upgrade corrective action. A Field Correction Acknowledgement Form and the Field Correction Completion Form was attached for customers to complete and return via the fax number 503-537-3617. The letter states that customers are to review their records in order to identify where the affected devices are located and which are still in service. Customers can call the Customer Service at 800-304-5332 and order the self-install Upgrade Kits with the Part Number 30792 at no charge. Customers are instructed to complete the Rework Correction Form once all of the lights have been updated and fax or mail to Marcus, attention Bonnie Dundas, at (503) 537-3617.
Quantity
13,226 units distributed worldwide
Distribution
Worldwide Distribution -- USA (nationwide) including states of: CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico and the countries of: Australia, Belgium, Canada, France, Germany, Ireland, Italy, Mexico, Norway, Poland, Russia, United Kingdom, Vietnam, and Trinidad.
Lot, serial or model codes
LuxStar Dental Light does not have Serial Number. They are identified by Catalog Model numbers such as ML1000, OL1000, CL1000, WL1000, CB1000, TL1000, and UL1000.
510(k) numbers
not on file
PMA numbers
not on file
Product code
EAZ
Firm FEI
1000523896
Firm city
Newberg
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register