FDA Medical Device Recalls (openFDA)
MicroAire Surgical Instruments, LLC: ASP-1020 LipoTower with Aspiration Pump. Medical use.
- Recall number
- Z-2821-2015
- Recalling firm
- MicroAire Surgical Instruments, LLC
- Product
- ASP-1020 LipoTower with Aspiration Pump. Medical use.
- Status
- Terminated
- Initiated
- 2015-06-12
- Posted
- 2015-09-30
- Terminated
- 2016-05-23
- Reason
- There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.
- Root cause
- Device Design
- Action
- The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.
- Quantity
- 3 LipoTower with Aspiration Pump
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Serial numbers 14031102, 15021105, and 14101103.
- 510(k) numbers
- ["K113128"]
- PMA numbers
- not on file
- Product code
- MUU
- Firm FEI
- 2020601
- Firm city
- Charlottesville
- Firm state
- VA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28