Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P

Recall number
Z-2790-2026
Recalling firm
Medline Industries, LP
Product
Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P
Status
Open, Classified
Initiated
2026-06-11
Posted
2026-07-23
Terminated
not on file
Reason
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Root cause
Process change control
Action
Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue with the device, potential harms, and requested the following: REQUIRED ACTION: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete the response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com 3. Upon receipt of the response form, the consignee account will receive over-labels to place on affected inventory, with instructions for the staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who transferred the affected product were directed to notify those to which the product was distributed or transferred. For contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
574 kits
Distribution
US: CT, FL, KS, OH, OR, NY, MD, MT, VA, MA.
Lot, serial or model codes
UDI/DI 10198459098109 (each) 40198459098100 (case), Lot Numbers: 25AMJ225, 25CMF838, 24KME164, 25DMB248, 25FMD941, 25GMB662, 25GMG296.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJG
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register