Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc.: Plum A+ Hyperbaric Infusion Pump. List Number 11005. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.

Recall number
Z-2722-2016
Recalling firm
Hospira Inc.
Product
Plum A+ Hyperbaric Infusion Pump. List Number 11005. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.
Status
Terminated
Initiated
2013-02-06
Posted
2016-09-06
Terminated
2016-11-14
Reason
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
Root cause
Nonconforming Material/Component
Action
Hospira sent an Urgent Device Field Correction letter, dated February 6, 2013, to customers. Customers were requested to return a reply form to Stericycle indicating that they have received the letter, and notified their users and/or customers. The customers were asked to supply the number of Plum A+ infusers at their facility. Product will be corrected by Hospira in the field.
Quantity
357,778 total
Distribution
United States nationwide distribution including PR, and USVI. Costa Rica, Turks and Caicos, and Dominican Republic.
Lot, serial or model codes
List Number 11005. Serial numbers: 16792201,  16792202,  16792205,  16792206,  16792208,  16792209,  16792210,  16792211,  16792213,  16792216,  16792217,  16792219,  16792220,  16792222,  16792223,  16792224,  16792225,  16792226,  16792227,  16792228,  16792230,  16792232,  16792233,  16792234,  16792235,  16792236,  16792238,  16792239,  16792240,  16792241,  16792242,  16792244,  16792245,  16792246,  16792248,  16792249,  16792250,  16792251,  16792252,  16792253,  16792254,  16792255,  16792256,  16792257,  16792258,  16792259,  16792260,  16792261,  16792263,  16792264,  16792265,  16792266,  16792268,  16792269,  16792271,  16792272,  16792273,  18807402,  18807405,  18807406,  18807407,  18807408,  18807409,  18807410,  18807413,  18807414,  18807417,  18807418,  18807420,  18807421,  18807423,  18807427,  18807428,  18807429,  18807430,  18807431,  18807432,  18807433,  18807434,  18807439,  18807443,  18807444,  18807446,  18807449,  18807452,  18807453,  18807455,  18807456,  18807459,  18807463,  18807466,  18807472,  18807474,  18807478,  18807484,  18807485,  18807486,  18807490,  18807491,  18807494,  18807496,  18807499,  18807500,  18807501,  18807506,  18807509,  18807510,  18807511,  18807513,  18807514,  18807518,  18807523,  18807524,  18807529,  18807535,  18807536,  18807539,  18807540,  18807541,  18807542,  18807543,  18807545,  18807546,  18807556,  18807557,  18807568,  18807574,  18807575,  18807576,  18807578,  18807581,  18807584,  18807585,  18807586,  18807588,  18807589,  18807590,  18807591,  18807596,  18807597,  18807598,  18807599,  18807600,  18807601,  18807602,  18807604,  18807605,  18807607,  18807608,  18807609,  18807610,  18807611,  18807612,  18807613,  18807614,  18807617,  18807618,  18807619,  18807620,  18807623,  18807625,  18807626,  18807627,  18807628,  18807630,  18807631,  18807632,  18807633,  18807634,  18807636,  18807637,  18807639,  18807640,  18807695,  18807696,  18807697,  18807698,  18807699,  18807700.
510(k) numbers
["K081412"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register