B. Braun Medical, Inc.: B.Braun Accuflo 100ML/H 100ML Elastomeric Infusion Device. AccuFlo devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications.
- Recall number
- Z-2655-2014
- Recalling firm
- B. Braun Medical, Inc.
- Product
- B.Braun Accuflo 100ML/H 100ML Elastomeric Infusion Device. AccuFlo devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications.
- Status
- Terminated
- Initiated
- 2014-08-06
- Posted
- 2014-09-16
- Terminated
- 2015-01-29
- Reason
- The primary packaging label (inner pouch label) indicates that the package contents are Catalog Item: CT-1000-250, while the secondary packaging label (outer carton label) indicates the contents are CT-1000-100. The actual contents are a CT-1000-100.
- Root cause
- Labeling Change Control
- Action
- B. Braun Medical Inc. provided the recall notification received from B. Braun Melsungen AG to Progressive Medical Inc. who is the initial importer of the product in the U.S. market. The notification was sent via FedEx and e-mail on/about August 6, 2014 and identified the product, problem, and action to be taken. The notification indicated that if the affected units were further distributed that any distributors are required to forward the letter on to their consignees.
- Quantity
- 8,640 units
- Distribution
- Distributed in the state of Missouri.
- Lot, serial or model codes
- Lot Number: 4A1125F111
- 510(k) numbers
- ["K081905"]
- PMA numbers
- not on file
- Product code
- MEB
- Firm FEI
- 2523676
- Firm city
- Allentown
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28