Skip to the content

FDA Medical Device Recalls (openFDA)

Medtest Holdings, Inc.: Pointe Scientific Liquid Creatine Kinase Reagent Set Device is a kit comprised of two reagents an R 1 Buffer and R2 CK enzyme solution. Intended Use: For the kinetic quantitative determination of creatine kinase activity in serum.

Recall number
Z-2650-2017
Recalling firm
Medtest Holdings, Inc.
Product
Pointe Scientific Liquid Creatine Kinase Reagent Set Device is a kit comprised of two reagents an R 1 Buffer and R2 CK enzyme solution. Intended Use: For the kinetic quantitative determination of creatine kinase activity in serum.
Status
Terminated
Initiated
2011-10-19
Posted
2017-06-29
Terminated
2017-07-24
Reason
Linear performance information in product insert does not match that listed in the approved premarket notification
Root cause
Labeling Change Control
Action
Pointe Scientific/Technical Bulletin sent notifications on October 19, /2011, TECHNICAL BULLETIN notifications were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, and consignee responsibilities.
Quantity
319.922 L
Distribution
Nationwide Distribution
Lot, serial or model codes
Models: C7522, HC422, HC722, HC922. Lot#'s: 020902; 029902 R1; 029902 R2; 029903; 113204; 129403; 209001. Exp dates 7/2012 to 9/2013.
510(k) numbers
K972155
PMA numbers
not on file
Product code
CGS
Firm FEI
1827821
Firm city
Canton
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28