Medtest Holdings, Inc.: Pointe Scientific Liquid Creatine Kinase Reagent Set Device is a kit comprised of two reagents an R 1 Buffer and R2 CK enzyme solution. Intended Use: For the kinetic quantitative determination of creatine kinase activity in serum.
- Recall number
- Z-2650-2017
- Recalling firm
- Medtest Holdings, Inc.
- Product
- Pointe Scientific Liquid Creatine Kinase Reagent Set Device is a kit comprised of two reagents an R 1 Buffer and R2 CK enzyme solution. Intended Use: For the kinetic quantitative determination of creatine kinase activity in serum.
- Status
- Terminated
- Initiated
- 2011-10-19
- Posted
- 2017-06-29
- Terminated
- 2017-07-24
- Reason
- Linear performance information in product insert does not match that listed in the approved premarket notification
- Root cause
- Labeling Change Control
- Action
- Pointe Scientific/Technical Bulletin sent notifications on October 19, /2011, TECHNICAL BULLETIN notifications were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, and consignee responsibilities.
- Quantity
- 319.922 L
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Models: C7522, HC422, HC722, HC922. Lot#'s: 020902; 029902 R1; 029902 R2; 029903; 113204; 129403; 209001. Exp dates 7/2012 to 9/2013.
- 510(k) numbers
- K972155
- PMA numbers
- not on file
- Product code
- CGS
- Firm FEI
- 1827821
- Firm city
- Canton
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28