Maquet Cardiovascular, LLC: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Recall number
Z-2636-2025
Recalling firm
Maquet Cardiovascular, LLC
Product
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
Status
Open, Classified
Initiated
2025-08-06
Posted
2025-09-25
Terminated
not on file
Reason
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Root cause
Under Investigation by firm
Action
Getinge Field Representatives contacted impacted accounts by telephone on 8/6/25 to notify them of this recall action and provide instructions for quarantine and return of affected devices. Consignees were asked to stop all visual inspection and return all HemoPro 3 units. Returns can be coordinated by calling or emailing Michele Kanis for an RMA at michele.kanis@getinge.com or (408)635-0647. Consignees are to utilize HemoPro 2 in place of HemoPro 3 units. Getinge Field Representatives delivered an updated URGENT MEDICAL DEVICE RECALL notice starting on 10/6/25 dated October 2025. The written notification is a supplement to the verbal notification made previously.
Quantity
98 units
Distribution
US Nationwide distribution in the states of FL, IL, MD & TX.
Lot, serial or model codes
Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.
510(k) numbers
["K251238"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register