Maquet Cardiovascular, LLC: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
- Recall number
- Z-2636-2025
- Recalling firm
- Maquet Cardiovascular, LLC
- Product
- Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
- Status
- Open, Classified
- Initiated
- 2025-08-06
- Posted
- 2025-09-25
- Terminated
- not on file
- Reason
- Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
- Root cause
- Under Investigation by firm
- Action
- Getinge Field Representatives contacted impacted accounts by telephone on 8/6/25 to notify them of this recall action and provide instructions for quarantine and return of affected devices. Consignees were asked to stop all visual inspection and return all HemoPro 3 units. Returns can be coordinated by calling or emailing Michele Kanis for an RMA at michele.kanis@getinge.com or (408)635-0647. Consignees are to utilize HemoPro 2 in place of HemoPro 3 units. Getinge Field Representatives delivered an updated URGENT MEDICAL DEVICE RECALL notice starting on 10/6/25 dated October 2025. The written notification is a supplement to the verbal notification made previously.
- Quantity
- 98 units
- Distribution
- US Nationwide distribution in the states of FL, IL, MD & TX.
- Lot, serial or model codes
- Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.
- 510(k) numbers
- ["K251238"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2242352
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28