FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal manipulators (arms) which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures in the pelvic, abdominal, and thoracic cavities.

Recall number
Z-2630-2017
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal manipulators (arms) which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures in the pelvic, abdominal, and thoracic cavities.
Status
Terminated
Initiated
2017-06-01
Posted
2017-06-06
Terminated
2017-10-02
Reason
da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.
Root cause
Process control
Action
Intuitive Surgical sent an Field Safety Notice letter dated June 1, 2017 to customers. The letter was sent electronically via through our da Vinci customer portal via email to the affected sites on June 1, 2017. Letters are sent to the da Vinci Coordinator, Recall Coordinator and/or Risk Management. The customer will receive a hard copy of this letter on June 6, 2017 via Fed Ex. The letter identified the affected product, problem and actions to be taken. For questions contact your Clinical Sales Representative.
Quantity
10 units
Distribution
Worldwide Distribution - US Nationwide in the states of TX, NY,FL, IL, OH, and countries of France, Ireland, Japan, and Taiwan
Lot, serial or model codes
Serial Numbers: SK0822, SK1154, SK1001, SK1170, SK1040, SK1176, SK1082, SK1180, SK1146, SK1190
510(k) numbers
["K131861"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register