Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M

Recall number
Z-2611-2026
Recalling firm
Medline Industries, LP
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M
Status
Open, Classified
Initiated
2026-05-28
Posted
2026-06-26
Terminated
not on file
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Root cause
Process change control
Action
On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
9,272 kits
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Lot, serial or model codes
DYNJ909090B UDI-DI 10195327652432 Lots 24EBL051 24GBG271 CDS982523M UDI-DI 10195327560591 Lots 24ALA227 24BLA369 24CLA108 24CLA518 24DLA082 24ELB053 24ILA112
510(k) numbers
K213481
PMA numbers
not on file
Product code
LRO
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register