FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
- Recall number
- Z-2610-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
- Status
- Open, Classified
- Initiated
- 2026-05-28
- Posted
- 2026-06-26
- Terminated
- not on file
- Reason
- Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
- Root cause
- Process change control
- Action
- On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 9,272 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
- Lot, serial or model codes
- DYNJ905503D UDI-DI 10193489430295 lots 21DBG611 21EBI605 21FBN751 DYNJ905503F UDI-DI 10193489967272 lots 21HBL060 21IBW859 22DBC340 22DBS437 22FBR267 22GBR286 22HBS560 22IBA577 22JBI883 22KBD587 22LBF425 23BBN903 23CBL450 23CBW776 23FBC209 23IBU343 24DBQ181 24FBT062 24GBR740 24HBI428 24JBP984 DYNJ905503G UDI-DI 10198459200397 lots 25ABJ069 25ABL749 25CBB762 25DBG799 25EBM994 25GBW226 25LBA950 25LBT019 26ABH519 26ABT315 26BBQ338 26DBJ110
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CAI
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28