FDA Medical Device Recalls (openFDA)
Computerized Medical Systems Inc: Focal Radiation Treatment Planning System, Focal Release 4.3.1 through 4.34.02. For use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.
- Recall number
- Z-2572-2011
- Recalling firm
- Computerized Medical Systems Inc
- Product
- Focal Radiation Treatment Planning System, Focal Release 4.3.1 through 4.34.02. For use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.
- Status
- Terminated
- Initiated
- 2008-03-31
- Posted
- 2011-06-15
- Terminated
- 2014-04-09
- Reason
- Focal: In absolute dose mode a beam with a port may be calculated with open normalization, with the result that the displayed dose reflects open normalization while the monitor units and QA plan reflect blocked normalization. This is particularly important when these beams also include compensators. The problem is detectable by looking at the delivered dose to the weight point in the patient as
- Root cause
- Software design
- Action
- Computerized Medical Systems Inc sent a "CUSTOMER ADVISORY" letter dated March 2008 to all affected customers. The letter describes the product, the problem and the actions to be taken. The firm provided a workaround for customers that states "Alternatively the plan can be imported directly into XIO". A copy of the Advisory was shipped with software within the affected range to those requesting a software upgrade. The issue was resolved in Focal Release 4.40.00 introduced in June 2008. The firm will send out a User Notice to all sites still running affected software to remind them that the issue resolution is available. For further questions, please call (403) 830-8023.
- Quantity
- 307
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of Albania, Argentina, Australia, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, France, Germany, Greece, India, Italy, Japan, Netherlands, Peru, Portugal, Qatar, Romania, Russia, Serbia, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, and Venezuela.
- Lot, serial or model codes
- Focal Release 4.3.1 through 4.34.02
- 510(k) numbers
- ["K013112"]
- PMA numbers
- not on file
- Product code
- MUJ
- Firm FEI
- 1937649
- Firm city
- Maryland Heights
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as