Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Information Technology, Inc.: MEDITECH Expanse Laboratory (LAB), Microbiology (MIC), Anatomical Pathology (PTH), Genetics (GEN). Calculator/data processing module for clinical use.

Recall number
Z-2569-2025
Recalling firm
Medical Information Technology, Inc.
Product
MEDITECH Expanse Laboratory (LAB), Microbiology (MIC), Anatomical Pathology (PTH), Genetics (GEN). Calculator/data processing module for clinical use.
Status
Open, Classified
Initiated
2025-02-05
Posted
2025-09-16
Terminated
not on file
Reason
Entering multiple keys that trigger input simultaneously may remove data from first field of screens and/or questionnaires with more than one field.
Root cause
Software design
Action
Consignees were notified via Salesforce email of this recall issue and provided instructions on workarounds
Quantity
148 units
Distribution
Worldwide distribution - US Nationwide and the countries of British Virgin Islands, Canada, Ireland & United Kingdom.
Lot, serial or model codes
Software Release: 2.1, 2.2 (using the Web Presentation Layer); UDI: 00863529000155.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
3009404844
Firm city
Canton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register