Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.

Recall number
Z-2564-2014
Recalling firm
Hitachi Medical Systems America Inc
Product
Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.
Status
Terminated
Initiated
2014-08-18
Posted
2014-09-02
Terminated
2015-09-14
Reason
The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
Root cause
Device Design
Action
Letters will be sent to customers. Hatachi Serivce will be sent to each site to exchange the Gradient Coil Assembly.
Quantity
39 systems
Distribution
Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, ID, KS, KY, MD, MS, NY, OH, and WY; countries of: Japan, Brazil, France, Germany, and Kyrgyzstan.
Lot, serial or model codes
Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information.
510(k) numbers
K113145
PMA numbers
not on file
Product code
LNH
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register