FDA Medical Device Recalls (openFDA)
Reshape Medical Inc: ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
- Recall number
- Z-2552-2016
- Recalling firm
- Reshape Medical Inc
- Product
- ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
- Status
- Terminated
- Initiated
- 2016-05-04
- Posted
- not on file
- Terminated
- 2016-08-17
- Reason
- Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure.
- Root cause
- Process control
- Action
- A Dear Doctor letter dated May 4, 2016, was sent to all the customers to inform them that ReShape Medical is recalling ReShape Integrated Dual Balloon System, Gen 1 due to a suspected misalignment of the connection between the proximal balloon cap and the delivery catheter. The letter informs the customers that the possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure. Customers are informed of the planned field correction and if customers have any questions then they're instructed to contact ReShape Medical Customer Service at (844) 937-7374. For further questions please call (949) 429-6680 ext. 106
- Quantity
- 56 units
- Distribution
- US: Distribution to the states of : NY, OH and CA.
- Lot, serial or model codes
- 160317-001, 160371-002, 160317-003, 160324-010.
- 510(k) numbers
- not on file
- PMA numbers
- P140012S002
- Product code
- LTI
- Firm FEI
- 3013508647
- Firm city
- San Clemente
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28