Skip to the content

FDA Medical Device Recalls (openFDA)

DiaSorin Molecular LLC: MOL3655 Simplexa VZV Swab Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA present in cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of VZV infection.

Recall number
Z-2549-2020
Recalling firm
DiaSorin Molecular LLC
Product
MOL3655 Simplexa VZV Swab Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA present in cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of VZV infection.
Status
Terminated
Initiated
2020-05-12
Posted
2020-07-10
Terminated
2024-04-19
Reason
Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results.
Root cause
Under Investigation by firm
Action
On 05/12/2020, the firm sent an "URGENT FIELD SAFETY NOTIFICATION LETTER" with an acknowledgement form to customers and distributors via email to inform them of an internal observed issue associated with the use of their Multi-Assay Suite (MAS) featured in the software version 2.0 where it may incorrectly apply the thresholds of the first loaded assay to the analytes coupled with the same dye in the second assay resulting in false negative or false positive results. The firm is instructing customers to: -Immediately discontinue the use of the MAS feature and that the issue will be correct in the new Software Version 2.1 which is expected to be release in June 2020. -Fill out the Response Form and return by either: -fax (562) 240-6526 Attention: Technical Services -email to: technical info.molecular@diasorin.com -mail to: ATTN: Technical Services, 11331 Valley View Street, Cypress, CA 90630 -Forward this communication to all required individuals with their organization. For further assistance with evaluating assay runs or for any questions, to contact Technical Services at +1-5672-240-6500, option 3 or via email ts.molecular@DiaSorin.com
Quantity
304 plastic vials (24 vials per box)
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, KS, MI, NC, PA, TX, and WA The countries of Austria, Czech Republic, France and Switerland.
Lot, serial or model codes
MOL3655 Simplexa VZV Swab Direct: Lot Numbers Release Date Exp Date 5743N 09/11/19 9/30/2020 5819N 10/09/19 10/31/2020 5820N 02/14/20 2/28/2021
510(k) numbers
["K192376"]
PMA numbers
not on file
Product code
PGI
Firm FEI
2023365
Firm city
Cypress
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register