FDA Medical Device Recalls (openFDA)
Zimmer Biomet, Inc.: Vanguard XP Tibial Tray 69 mm Item # 195248
- Recall number
- Z-2531-2019
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Vanguard XP Tibial Tray 69 mm Item # 195248
- Status
- Terminated
- Initiated
- 2019-08-15
- Posted
- not on file
- Terminated
- 2021-04-16
- Reason
- The locking bar not fully engaging
- Root cause
- Manufacturing material removal
- Action
- Risk Manager Responsibilities: 1.Review this notification and ensure that affected personnel are aware of the contents. 2.If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1.Review this notification for awareness of the contents. 2.There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3.A visual and audible confirmation should be made to ensure complete locking bar insertion. 4.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 5.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-
- Quantity
- 123 units
- Distribution
- State NY IN OH MI GA SC FL MO WI MN ND SD NE NJ PA MD IL IA CA VA NC DC WV AR AL MS AL MI WA OR ID LA FL KS TN MA KY OK TX TX CT AZ NV NM MT WY CO UT DC WY Country: KOREA CHINA CANADA AUSTRALIA JAPAN EMEA SWEDEN ITALY
- Lot, serial or model codes
- Lot Number 117720 332660 373010 373000 514970 515000 372990 039210 112980 113000 112990 202920 303370 303380 590480 660400 660410 584210 590490 833180 752040 752050 752560 752060 919220 928860 754470 764360 837750 764350 846490 861180 881480 881490 928870 846500 881460 984370 984380 984360 837760 570940
- 510(k) numbers
- ["K122160"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28