Skip to the content

FDA Medical Device Recalls (openFDA)

LEICA BIOSYSTEMS NUSSLOCH GMBH: Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.

Recall number
Z-2529-2024
Recalling firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Product
Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.
Status
Open, Classified
Initiated
2024-05-10
Posted
2024-08-07
Terminated
not on file
Reason
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
Root cause
Device Design
Action
The recalling firm issued letters dated 4/30/2024 via email on 5/10/2024. The letter provided the model and serial number of the devices involved, a description of the problem, and immediate actions to be taken. The consignee is to check to see if they have one of the affected instruments and determine whether the Arcadia H is in active use in the facility. If it is, the consignee is to discontinue the use of it immediately and disconnect it from main voltage. If it has not been decommissioned at the time of receipt of the letter, the consignee is to reach out to their sales representative. The sales representative will work with the consignee to determine the disposition of the device according to their local regulations. The consignee is requested to pass the notification letter to users of the device and to all within their organization who need to know. The consignee is to confirm receipt of the letter within 5 days by completing the attached acknowledgement form and returning a scanned copy via email.
Quantity
180 devices
Distribution
US Nationwide, including government distribution.
Lot, serial or model codes
Model/Catalog Number: 14039354100; UDI-DI: 04049188194763, Serial numbers: 58, 62, 98, 99, 101, 117,120, 121, 122, 123, 124, 125, 127, 128, 129, 177, 178, 179, 180, 181, 185, 186, 218, 222, 223, 279, 282, 284, 286, 287, 288, 289, 290, 291, 292, 314, 315, 317, 318, 319, 321, 322, 353, 354, 355, 356, 357, 358, 360, 363, 365, 366, 371, 478, 480, 483, 484, 487, 488, 489, 490, 491, 492, 553, 560, 561, 562, 584, 586, 588, 589, 590, 593, 597, 643, 644, 647, 649, 650, 651, 652, 653, 656, 658, 659, 697, 698, 705, 710, 714, 715, 716, 717, 739, 740, 741, 742, 743, 757, 819, 820, 821, 822, 825, 826, 827, 828, 829, 830, 831, 832, 878, 879, 880, 881, 882, 883, 884, 885, 886, 887, 888, 889, 890, 891, 892, 893, 894, 896, 897, 898, 900, 901, 904, 906, 907, 909, 911, 919, 921, 922, 923, 924, 927, 928, 930, 1113, 1114, 1115, 1116, 1117, 1118, 1123, 1124, 1198, 1199, 1200, 1201, 1202, 1203, 1204, 1205, 1206, 1207, 1212, 1216, 1217, 1266, 1267, 1268, 1271, 1273, 1274, 1277, 1278, 1279, 1280, 1283, and 1285. Model/Catalog Number: 14039357258; UDI-DI: 04049188205971, serial number 5033.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IDW
Firm FEI
3002179321
Firm city
Nussloch
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register