GE Medical Systems, LLC: MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE Signa 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution high signal-to-noise ratio, with short scan times. The Signa Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and tile surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.

Recall number
Z-2527-2018
Recalling firm
GE Medical Systems, LLC
Product
MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE Signa 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution high signal-to-noise ratio, with short scan times. The Signa Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and tile surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.
Status
Terminated
Initiated
2018-03-30
Posted
not on file
Terminated
2024-03-29
Reason
There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.
Root cause
Nonconforming Material/Component
Action
GE Healthcare notified customers on about 03/30/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter informed customers that the MR Surgical Suited System could still remain in use, it referred them to the GE MR Surgical Suite Operator's Manual for detailed descriptions of the various safety interlocks and operating mechanisms, encouraged users to monitor for any issues releasing the transfer board from the GE MR Surgical Table, and instructed users to test the patient transfer features as part of a quality check prior to each use. The letter also informed customers that a GE Healthcare representative will contact customers to ensure they have the latest operator manual at their site, update preventative maintenance documentation to include a procedure that focuses specifically on this equipment, and GE Healthcare representative will contact customers for scheduling a visit at which time GEHC will conduct this preventative maintenance procedure for their device and install new hardware. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your the Service Representative.
Quantity
54 units total
Distribution
US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Model Numbers M8074SS, M8053SS, M0074SS, M0001SS, 5265300-7, 5265300-8.   Serial Number (System ID Number): 00000000000002 (602406IMR), 4295404 (4295404MR11), N/A (415UCBHMR3), N/A (720848MR4), 00000304124MR1 (904202MR8), N/A (309655IMRI), 4425007 (913588NMRW), 00000000UA0117 (616267IMRI), 00000000000003 (507284NMRV), 4311497 (MR229054),  4510086 (212636SIGNA3T), 00000000UA0116 (330344MR750W), UA0326 (614293MR9), N/A (614RMHTMR1), 4371613 (412692MR750W), N/A (214820ITABLE), 4250114 (409772MR6), 4550641 (4550641MR11), 4482037 (757388750WMR), 00000012228M63 (608263MR4IOR), 4356151 (414805IMR1), 00000299229MR5 (AH5824MR01), 00000012051M69 (AH5824SS01), 00000021158M61 (MRHM0885), 00000014131M67 (MRHM0323), 00000018480M64 (MR307144MR6), 4565719 (082427110075),  00000013254M68 (082445160014), 000000PTST0008 (082427040101),  000000PTST0007 (082427050024), 4596631 (082427310113), 4409769 (082427140133), 000000PTST0003 (CS1020MR01), DUMFMI60885001 (CS1006MR03), 00001309FMM07L (T4185506), 00000021695M62 (M4194477), 00000021695M62 (X41944020), 00000305145MR5 (UC2547MR01), 4320023 (083027822002316), 4508956 (A5381507), 00000010003M62 (EM0293), 00001219FMM0H1 (EM0105), 00000017217M61 (EM0105), 000000PTST0010 (YM3833), 00000000000004 (YM1733),  2937140 (0850270414), 4548529 (KW1008MR08), 000000PTST0006 (ZA2169MR01), 00001251FMM044 (34368MRS01), 4344097 (34418MRS01),  4605737 (N/A), 000000PTST0005 (10590MRS01), DUMFMI60885002 (10590MRS02), 4569734 (00132MRS01)
510(k) numbers
["K041476"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2183553
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register