Skip to the content

FDA Medical Device Recalls (openFDA)

PFM MEDICAL INC.: Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

Recall number
Z-2526-2026
Recalling firm
PFM MEDICAL INC.
Product
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Status
Open, Classified
Initiated
2026-04-21
Posted
2026-06-23
Terminated
not on file
Reason
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Root cause
Under Investigation by firm
Action
On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to recall@pfmmedicalusa.com
Quantity
32
Distribution
US Nationwide distribution in the states of NY, MD.
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)
510(k) numbers
["K011783"]
PMA numbers
not on file
Product code
DXE
Firm FEI
3013666218
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register