Intuitive Surgical, Inc.: Da Vinci ASSY, PSS, SP1098, Part Number: 380601

Recall number
Z-2521-2026
Recalling firm
Intuitive Surgical, Inc.
Product
Da Vinci ASSY, PSS, SP1098, Part Number: 380601
Status
Open, Classified
Initiated
2026-04-22
Posted
2026-06-22
Terminated
not on file
Reason
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
Root cause
Process control
Action
On 4/22/2026, correction notices were mailed and emailed to customers who were asked to do the following: 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Please retain a copy of this letter, place a copy with your affected system, and keep a copy of the acknowledgement form for your files. 3. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. A firm Representative will schedule a site visit to perform an inspection of the system. If you need further information or support concerning this Medical Device Correction, please contact your da Vinci SP Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Quantity
454
Distribution
Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Greece, Taiwan, Australia, China, Turkey, Poland, United Arab Emirates, Hong Kong, Romania.
Lot, serial or model codes
UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 944585, 10065786, 935214, 10020729, 10012963, 10020705, 10031145, 10031144, 973825, 10012962, 10036960, 963856, 10711066, 10035476, 795038, 10043635, 944583, 10072852, 10072853, 966079, 10069899, 10076490, 10087446, 10054478, 10080929, 954096, 10130089, 10165287, 942811, 10180162, 10172589, 10103483, 10103484, 10056518, 10092007, 10246370, 10209961, 10183740, 10232634, 504007, 10231116, 10211880, 10194911, 10245457, 10112470, 10189443, 977044, 10238314, 10644376, 10036961, 10274101, 10112471, 10238315, 10172574, 10281524, 10278366, 10252751, 10289629, 10293276, 10261900, 10274102, 10285085, 10281525, 10243796, 10267796, 10278365, 10285084, 10308532, 10318311, 10293274, 10323177, 10322512, 10289630, 10326939, 10293275, 10333333, 10326938, 10492299, 10302474, 10302473, 10368392, 10332310, 10346946, 10354749, 10346943, 10333332, 10346942, 10267795, 10323178, 10295600, 10223509, 10252780, 10396330, 10263730, 10354748, 10337879, 10265272, 10354750, 10384455, 10359960, 10368393, 10366206, 10375031, 10363897, 10396329, 10363898, 10375032, 10375033, 10409934, 10401560, 10409933, 10408268, 10294693, 10408269, 10396331, 10480883, 10499186, 10499185, 10384456, 10502598, 10495654, 10502599, 10505838, 10512168, 10512167, 10512166, 10508290, 10368391, 10556776, 10417803, 10526733, 10417164, 10508289, 10384454, 10492298, 10480878, 10526735, 10526736, 10537191, 10495655, 10526734, 10526732, 530114, 10537202, 10568118, 10308533, 10547839, 10564979, 10550219, 10564978, 10550220, 10550221, 10561240, 10594744, 10575228, 10571615, 10575229, 10583123, 10571614, 10556769, 10583124, 10594746, 10556777, 10564977, 10594743, 10579531, 10613173, 10568117, 10571616, 10615325, 10583126, 10619701, 10594742, 10541679, 10579530, 10561239, 10594745, 10615326, 10590323, 10590322, 10583125, 10615941, 10608089, 10644375, 10604955, 10561241, 10624252, 10647203, 10670710, 10672825, 10624990, 10295599, 10633489, 10622883, 10647201, 10624251, 10686903, 10612973, 10619702, 10672839, 10696000, 10615940, 10659191, 10651445, 10651443, 10682596, 10702363, 10644374, 10653815, 10651444, 10627633, 10686902, 10672840, 10627632, 10695999, 10665125, 10604956, 10480877, 10682597, 10694981, 10633487, 10716881, 10659189, 10728578, 10627631, 10702364, 10734551, 10659190, 10633488, 10734552, 10706206, 650202, 10744259, 10706207, 10766233, 10766236, 10793380, 10759158, 10759159, 10759161, 10763608, 10763609, 10808841, 10766234, 10763611, 10741377, 10759160, 10763610, 10826493, 10776325, 10766235, 10741378, 10744258, 10734553, 10808839, 10808840, 10771126, 10803470, 10793379, 10817400, 10798922, 10803484, 10766232, 10887260, 10830719, 10771125, 10776323, 10815457, 10744257, 10823000, 10776324, 10782092, 10873054, 10879012, 10897582, 10856718, 10889592, 10864698, 10922820, 10912617, 10647204, 10869722, 10889594, 10871773, 10845441, 10859690, 10826495, 10579532, 10782090, 10826494, 10792831, 10883203, 10871774, 10856716, 10856732, 10859689, 10947466, 10798915, 10968849, 10952495, 10957321, 10875232, 10903586, 10927370, 10947464, 10951328, 10981839, 10787962, 10908841, 10908842, 10817401, 10938311, 10933815, 10938312, 10955031, 10927371, 10933816, 10998396, 10927372, 10240354, 10998395, 11004904, 10979952, 10875231, 10912618, 10897581, 10938310, 10991256, 10991257, 10991259, 10895278, 10883202, 10889593, 10912619, 11016764, 11013103
510(k) numbers
K173906, K182371
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register