FDA Medical Device Recalls (openFDA)
Medline Industries, LP: MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.
- Recall number
- Z-2488-2026
- Recalling firm
- Medline Industries, LP
- Product
- MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.
- Status
- Open, Classified
- Initiated
- 2026-05-04
- Posted
- 2026-06-17
- Terminated
- not on file
- Reason
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Root cause
- Nonconforming Material/Component
- Action
- Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 1248 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Lot, serial or model codes
- Medline Kit Number/SKU DYNJ912271: UDI/DI 10198459688324 (each), 40198459688325 (case), Lot Number 26CBQ109; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 25FBO720; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO387; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO015; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBD819; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBA736; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24FBG144; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24DBO838; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24BBC453; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24ABM118; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23KBS950; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23IBC670; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22LBN883; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22KBF121; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22IBN293; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22FBO695; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22OBH349; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBY498; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBY507; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBH105; Medline Kit Number/SKU DYNJ905368D: UDI/DI 10198459698149 (each), 40198459698140 (case), Lot Number 26DBE022; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25GBD045; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25EBN367; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25CBR397; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25BBK435; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25ABR151; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24JBU152; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24IBD106; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24HBI890; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24GBS879; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24EBB182; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24CBT638; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24CBN624; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23KBU577; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23IBS119; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23GBV028; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23HBO351; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23BBL739; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23BBL740; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23ABS355; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23ABG430; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG375; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG376; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG377; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22EBM264; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22EBL230; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 21LBV309; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21IMB152; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21HMD185; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21DMA957; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21BME627; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21BMC286.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OFF
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28