Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Clinical Diagnostics: VITROS 3600 Immunodiagnostic System - Refurbished Product Code: 6802914 Unique Identifier: 10758750007103

Recall number
Z-2481-2020
Recalling firm
Ortho Clinical Diagnostics
Product
VITROS 3600 Immunodiagnostic System - Refurbished Product Code: 6802914 Unique Identifier: 10758750007103
Status
Terminated
Initiated
2020-05-20
Posted
not on file
Terminated
2022-11-16
Reason
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances
Root cause
Under Investigation by firm
Action
Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
Quantity
27 ( U.S.= 0; OUS= 27)
Distribution
US Nationwide Distribution Foreign: Country AR AU BR CA CL CN CO DO FR ID IN JP KR MX MY NP PE PH SG TH TT VE VN
Lot, serial or model codes
Affected Serial Number : 36000192 36000396 36000430 36000431 36000446 36000468 36000504 36000517 36000577 36000580 36000585 36000631 36000647 36000366 36000267 36000386 36000404 36000420 36000562 36000614 36000575 36000589 36000480 36000582 36000687 36000492 36001042
510(k) numbers
["K083173"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register