Alphatec Spine, Inc.: LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF

Recall number
Z-2475-2023
Recalling firm
Alphatec Spine, Inc.
Product
LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,
Status
Open, Classified
Initiated
2023-07-07
Posted
2023-08-28
Terminated
not on file
Reason
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.
Root cause
Nonconforming Material/Component
Action
On 7/7/23, recall notices were emailed to consignees who were asked to do the following: 1) Abstain from sale and/or use and return to the recalling firm. 2) Complete and return the response form via email to QMSCompliance@atecspine.com Questions/concerns can be directed to 1-800-922-1356 or QMSCompliance@atecspine.com
Quantity
29
Distribution
US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.
Lot, serial or model codes
UDI-DI: 00190376228037, Lot: EM50715
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWJ
Firm FEI
2027467
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source