Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard Inc: Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF: 119314; SureStep Foley Tray System Lubri-Sil I.C. Complete Care, REF: A319514A; Bard Silver Lubri-Sil Foley Tray Temperature Sensing with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29000J14, 29030J14, 29000J14S; Bard Silver Lubri-Sil Foley Tray Temperature Sensing, Complete Care with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29030J14S

Recall number
Z-2462-2025
Recalling firm
C.R. Bard Inc
Product
Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF: 119314; SureStep Foley Tray System Lubri-Sil I.C. Complete Care, REF: A319514A; Bard Silver Lubri-Sil Foley Tray Temperature Sensing with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29000J14, 29030J14, 29000J14S; Bard Silver Lubri-Sil Foley Tray Temperature Sensing, Complete Care with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29030J14S
Status
Open, Classified
Initiated
2025-06-18
Posted
2025-08-28
Terminated
not on file
Reason
Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.
Root cause
Process design
Action
On 6/18/2025, recall notices were mailed or emailed to Recall Coordinators, Directors of Nursing, Purchasing, Risk Management, and distributors who were informed the following: The affected product was distributed from 07 March 2024 to 14 May 2025. - If affected product is currently indwelling and draining urine appropriately, the catheter does not need to be removed. - For affected product currently not in use, quarantine, and dispose of unused product and source alternative product. - Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Recall (Correction). - If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. - Complete and return the attached Customer Response Form via Email BDRC26@bd.com Report any complaints experienced with the use of this product to the firm via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) say "Product Complaints" when prompted Mon-Fri 8:00am and 5:00pm CT, Email: productcomplaints@bd.com Firm's Medical Information Services: 1-800-555-7422, medical.information@bd.com
Quantity
175,158
Distribution
Worldwide - US Nationwide distribution in the states of IA, PA, AR, OR, MN, IN, WV, OK, AL, MD, HI, GA, AZ, MA, FL, CA and the countries of Japan.
Lot, serial or model codes
Catalog/UDI-DI/Lot(Expiration): 119314/00801741039843, 10801741039840/NGJS3588(31/03/2029), NGKQ2500(21/12/2029), NGKQ0170(30/11/2029), NGJW2499(31/08/2029), NGJT1284 (31/03/2029), NGKQ2492(31/12/2029), NGKP3152(31/12/2029), NGJX0215(31/07/2029). A319514A/00801741100321, 10801741100328/NGJN0498(31/01/2026), NGJW3127(30/04/2027) 29000J14/21200BZY00110000/MYJX0683(28/05/2027), MYJP0751(28/02/2027), MYJR1984(28/04/2027), MYJQ0070(28/04/2027), MYJS6278(28/04/2027), MYJR3026(28/04/2027), MYJT3766(28/04/2027), MYJY6197(28/06/2027), MYJX2833(28/06/2027), MYJQ4022(28/04/2027), MYJT3764(28/04/2027), MYJZ0962(28/06/2027), MYJS4679(28/04/2027), MYJX0686(28/06/2027), MYJV4359(28/03/2027), MYJS3597(28/04/2027), MYJW1900(28/04/2027), MYJT3765(28/04/2027), MYJY1245(28/06/2027), MYJY1242(28/06/2027), MYJY1248(28/06/2027). 29030J14/21200BZY00110000/MYJX2846(28/06/2027), MYJP2013(28/02/2027), MYJT1017(28/04/2027), MYJW1906(28/04/2027), MYJT4535(28/07/2027), MYJP4407(28/04/2027), MYJS1895(28/04/2027), MYJW1904(28/04/2027), MYJV4102(28/03/2027), MYJX2836(28/05/2027), MYJZ4526(28/06/2027), MYJT3468(28/04/2027), MYJP2012(28/02/2027), MYJV4099(28/03/2027), MYJZ0964(28/06/2027), MYJQ0071(28/04/2027), MYJS6279(28/04/2027), MYJY4511(28/06/2027), MYJQ4024(28/04/2027), MYJZ0963(28/06/2027), MYJT1018(28/04/2027), MYJR3027(28/04/2027), MYJS3599(28/04/2027), MYJQ4023(28/04/2027), MYJR3083(28/04/2027), MYJY4512(28/06/2027), MYJQ0073(28/04/2027), MYJS1893(28/04/2027), MYJN2580(28/01/2027), MYJU2705(28/08/2027), MYJZ4527(28/06/2027), MYJS1894(28/04/2027), MYJV4103(28/03/2027), MYJR3059(28/04/2027), MYJU0772(28/08/2027), MYJV4101(28/03/2027), MYJY4513(28/06/2027), MYJQ0072(28/04/2027), MYJU3846(28/03/2027), MYJU3847(28/05/2027), MYJX2841(28/06/2027), MYJX2840(28/05/2027), MYJX2839(28/05/2027), MYJY4510(28/06/2027), MYJY4509(28/06/2027), MYJV4100(28/06/2027), MYJU3848(28/06/2027). 29000J14S/21200BZY00110000/MYJY1249(28/06/2027), MYJY0937(28/05/2027), MYJW1903(28/03/2027). 29030J14S/21200BZY00110000/MYJU0773(28/08/2027)
510(k) numbers
["K070582"]
PMA numbers
not on file
Product code
EZL
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register