Skip to the content

FDA Medical Device Recalls (openFDA)

Omnilife Science Inc.: APEX INTERFACE Hemi 54X and Flared 56X, 3 Hole Acetabular Shells; Model: H5-12354, H5-11356. Product Usage: The Apex Interface Acetabular System is intended for use in combination with the Apex Modular Hip Stem in total hip replacement procedures.

Recall number
Z-2445-2015
Recalling firm
Omnilife Science Inc.
Product
APEX INTERFACE Hemi 54X and Flared 56X, 3 Hole Acetabular Shells; Model: H5-12354, H5-11356. Product Usage: The Apex Interface Acetabular System is intended for use in combination with the Apex Modular Hip Stem in total hip replacement procedures.
Status
Terminated
Initiated
2015-06-04
Posted
2015-08-21
Terminated
2015-09-28
Reason
The device may have improper screw hole placement due to inaccurate location of the index line.
Root cause
Process design
Action
Omni sent an Urgent Medical Device Field Correction letter dated June 4, 2015 via email to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review inventory for affected product and complete and return the attached Acknowledgement and Receipt Form by FAX to 508-819-3390.
Quantity
112
Distribution
US Nationwide Distribution in the states of IL, VA, CO, UT, NY, FL, MA, OK, CA, TN and WY.
Lot, serial or model codes
H5-12354: Lots - 20127 and 20167; H5-11356: Lot - 20097
510(k) numbers
K112779
PMA numbers
not on file
Product code
LPH
Firm FEI
3008342029
Firm city
East Taunton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register