Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics, Inc.: Solero Generator US PG, Item No. H78712740000US0, Catalog No. 12740000US

Recall number
Z-2437-2021
Recalling firm
Angiodynamics, Inc.
Product
Solero Generator US PG, Item No. H78712740000US0, Catalog No. 12740000US
Status
Terminated
Initiated
2021-07-20
Posted
not on file
Terminated
2023-03-15
Reason
Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up.
Root cause
Software design
Action
On July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees. Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices. In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere to the following Warning in the current product labeling (User Manual): WARNING: Do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "Ready".
Quantity
26 US, 4 OUS
Distribution
Domestic distribution Nationwide. Foreign distribution worldwide.
Lot, serial or model codes
Serial Nos. QBY0002693 QBY0002376 QBY0002765 QBY0002380 QBY0002763 QBY0002662 QBY0002570 QBY0002683 QBY0002551 QBY0002618 QBY0002006 QBY0002293 QBY0002578 QBY0002666 QBY0002378 QBY0002375 QBY0002283 QBY0002539 QBY0002762 QBY0002681 QBY0002690 QBY0002389 QBY0002518 QBY0002670 QBY0002395 QBY0002687 QBY0002613 QBY0002664 QBY0002759 QBY0002780
510(k) numbers
["K162449"]
PMA numbers
not on file
Product code
NEY
Firm FEI
1319211
Firm city
Queensbury
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register