Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp.: Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85

Recall number
Z-2434-2023
Recalling firm
Datascope Corp.
Product
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85,
Status
Open, Classified
Initiated
2023-07-31
Posted
2023-08-30
Terminated
not on file
Reason
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
Root cause
Device Design
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. System Over-Temperature User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. When a System Over Temperature alarm is triggered, the device must be powered down andallowed to cool to a safe operating internal temperature before therapy can be re-initiated using the same IABP.Should you experience a System Over Temperature of Cardiosave IABP during therapy, perform the following instructions documented in CARDIOSAVE Hybrid and Rescue, Operational Instruction for US and OUS: 1. Turn the IABP OFF by pressing and holding the green IABP Power Button, located on theback panel, for 2 seconds 2. Wait 10 seconds 3. Turn the IABP ON by pressing and releasing the green IABP Power Button If the alarm message persists, switch to another IABP if available and contact your Service representative. If another IABP console is not available for use, alternative means of providing hemodynamic support (vasopressors, inotropes or alternate therapies) may be initiated by a healthcare provider as a temporizing measure. Consideration should be given to having a backup IABP available in the event the IABP doesn t cool sufficiently to restart or the alarm reoccurs. To help prevent such alarms from occurring, ensure that there are no restrictions to the airflow around the Cardiosave device. Restrictions to the vents on the device can significantly increase the internal temperature of the device, leading to such temperature excursions. Additionally, as per CARDIOSAVE Hybrid and Rescue Operational Instruction, it is essential not to operate the Cardiosave device outside of the posted operating ambient ranges. This is to prevent such temperature excursions and ensure the safe and effe
Quantity
47 units (42 US, 5 OUS)
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449, Model: 0998-UC-0800-75, UDI: N/A; Model: 0998-UC-0800-83, UDI: N/A; Model: 0998-UC-0800-85, UDI: N/A;
510(k) numbers
K181122
PMA numbers
not on file
Product code
DSP
Firm FEI
3001418283
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source