FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments.
- Recall number
- Z-2423-2010
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments.
- Status
- Terminated
- Initiated
- 2010-05-05
- Posted
- 2010-09-15
- Terminated
- 2010-10-25
- Reason
- Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.
- Root cause
- Nonconforming Material/Component
- Action
- Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units.
- Quantity
- 228 units
- Distribution
- Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland.
- Lot, serial or model codes
- Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments within the serial number range of B01060900055 through B01090100414R0, excluding B01081100388R0, B01081100389R0, B01081200396R0, B01081200402R0, B01081200403R0, B01081200404R0, B01081200406R0, B01081200408R0, B01081200410R0, B01081200411R0, B01090100412R0, B01090100413R0.
- 510(k) numbers
- K051087
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28