Alcon Research LLC: Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
- Recall number
- Z-2412-2024
- Recalling firm
- Alcon Research LLC
- Product
- Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
- Status
- Open, Classified
- Initiated
- 2024-05-23
- Posted
- 2024-07-18
- Terminated
- not on file
- Reason
- After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
- Root cause
- Software design
- Action
- On 5/23/24, correction notices were mailed to customers informing them of the following: Prior to ending procedure, re-perform registration to ensure proper axis alignment. Firm is expediting a software update that is expected to be released to upgrade systems running on software version 1.5 starting mid-June 2024. Firm's Technical Service representative will contact you to schedule your service appointment for this software upgrade. If you choose to use the image guidance functionality for toric alignment, to mitigate the potential for this anomaly, users are advised to re-perform registration prior to ending the case or if any axis discrepancy is noted during the case. To re-perform registration, go to the registration step, re-confirm the reference image to the new surgery image, adjust the registration angle (if necessary), select "Confirm" and proceed to toric alignment or desired step. Complete and return the response form via email to Market.Actions@Alcon.com If you have any questions regarding this issue, please contact your Firm representative or Technical Support at 1-800-832-7827 or contact your local Sales Representative.
- Quantity
- 230
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
- Lot, serial or model codes
- UDI-DI: 00380650003997. Software Version: 1.5. Serial Numbers: 004-0680001, 004-0680002, 004-0680003, 004-0680004, 004-0680007, 004-0680008, 004-0680017, 004-0680026, 004-0680036, 004-0680047, 004-0680048, 004-0680086, 004-0680090, 004-0680092, 004-0680096, 004-0680100, 004-0680107, 004-0680112, 004-0680120 004-0680127,004-0680131, 004-0680134, 004-0680140, 004-0680152, 004-0680153, 004-0680161, 004-0680163, 004-0680188, 004-0680190,004-0680203, 004-0680206, 004-0680266, 004-0680267, 004-0680271, 004-0680273, 004-0680274, 004-0680283, 004-0680288, 004-0680293, 004-0680294, 004-0680306, 004-0680311, 004-0680315, 004-0680319, 004-0680325, 004-0680334, 004-0680336, 004-0680337, 004-0680338, 004-0680339, 004-0680346, 004-0680348, 004-0680352, 004-0680362, 004-0680404, 004-0680417, 004-0680418, 005-7230003, 005-7230005, 005-7230007, 005-7230017, 005-7230019, 005-7230021, 005-7230022, 005-7230023, 005-7230031, 005-7230035, 005-7230057, 005-7230081, 005-7230085, 005-7230086, 005-7230102, 005-7230108, 005-7230109, 005-7230186, 007-4610018, 007-4610023, 007-4610030, 007-4610032, 007-4610033, 007-4610034, 007-4610037, 007-4610044, 007-4610074, 007-4610107, 007-4610112, 007-4610115, 007-4610122, 007-4610123, 007-4610124, 007-4610125, 007-4610127, 007-4610132, 007-4610138, 007-4610166, 007-4610167, 007-4610179, 007-4610181, 007-4610186, 007-4610217, 007-4610219, 007-4610220, 007-4610226, 007-4610229, 007-4610230, 007-4610262, 007-4610275, 007-4610283, 007-4610287, 007-4610288, 007-4610308, 007-4610309, 007-4610351, 007-4610355, 007-4610359, 007-4610360, 007-4610366, 007-4610394, 007-4610404, 007-4610405, 007-4610406, 007-4610409, 009-4570001, 009-4570004, 009-4570005, 009-4570007, 009-4570013, 009-4570014, 009-4570015, 009-4570016, 009-4570019, 009-4570022, 009-4570033, 009-4570035, 009-4570044, 009-4570047, 009-4570048, 009-4570049, 009-4570051,009-4570052, 009-4570053 009-4570054,009-4570057, 009-4570059, 009-4570060, 009-4570064, 009-4570068, 009-4570070, 009-4570072, 009-4570073, 009-4570096, 009-4570097, 009-4570099, 009-4570100, 009-4570114, 009-4570116, 009-4570125, 009-4570127, 009-4570129, 009-4570130, 009-4570136, 009-4570137, 009-4570144, 009-4570148, 009-4570149, 009-4570150, 009-4570153, 009-4570155, 009-4570156, 009-4570158, 009-4570160, 009-4570161, 009-4570167, 009-4570171, 009-4570173, 009-4570176,009-4570178, 009-4570179, 009-4570180, 009-4570183, 009-4570184, 009-4570200, 009-4570204, 009-4570207, 009-4570233, 009-4570249, 009-4570256, 009-4570259, 009-4570263, 009-4570270, 009-4570275, 009-4570278, 009-4570281, 009-4570289, 009-4570293, 009-4570294, 009-4570300, 009-4570308, 009-4570317, 009-4570320, 009-4570331, 009-4570342, 009-4570343, 009-4570350, 009-4570353, 009-4570357, 009-4570365, 009-4570368, 009-4570393, 009-4570397, 009-4570425, 009-4570431, 009-4570435, 009-4570440, 009-4570448, 009-4570449, 009-4570451, 009-4570455, 009-4570457, 009-4570459, 009-4570460, 009-4570474, 009-4570489, 009-4570508, 009-4570510, DM550021, DM550024, DM550043, DM550051, DM550054
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FWJ
- Firm FEI
- 1610287
- Firm city
- Fort Worth
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28