Skip to the content

FDA Medical Device Recalls (openFDA)

Whatman, Inc.: Whatman Body Fluid Collection Paper- 903 California Newborn Screening Card Ref 110534 903 Card are intended for use by healthcare professionals in the collection and in-vitro storage of neonate human blood. The neonate blood is tested to screen the infant for congenital abnormalities of metabolism and other conditions.790

Recall number
Z-2384-2010
Recalling firm
Whatman, Inc.
Product
Whatman Body Fluid Collection Paper- 903 California Newborn Screening Card Ref 110534 903 Card are intended for use by healthcare professionals in the collection and in-vitro storage of neonate human blood. The neonate blood is tested to screen the infant for congenital abnormalities of metabolism and other conditions.790
Status
Terminated
Initiated
2009-10-29
Posted
2010-09-07
Terminated
2020-08-24
Reason
Neonate newborn screening card demographic and filter portion may be detached
Root cause
Process control
Action
The firm, Whatman, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 29, 2009, to the customer. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: inspect the card and ensure the 903 Paper Portion is firmly and completely fixed to the Demographic portio, prior to recording patient demographic details on the card and obtaining a patient sample; discard the card, obtain a new card and inspect as above, if the 903 Paper Portion of the card partially or completely detaches from the Demographic portion of the card; and to please staple the card stub at two points as specified in the illustration, to prevent any opportunity for 903 paper detachment from the demographic portions. If Whatman customers/users have any questions in relation to this customer notification, please contact, either Product Manager at +1 732 885 6570,email: lili.lee@ge.com or Quality Assurance at +1 207 459 7557, email: heidi.kuell@ge.com.
Quantity
488,000 cards
Distribution
Nationwide distribution: CA
Lot, serial or model codes
Lot: W 071 6820708
510(k) numbers
["K932661"]
PMA numbers
not on file
Product code
JKA
Firm FEI
1281317
Firm city
Sanford
Firm state
ME
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register