FDA Medical Device Recalls (openFDA)
Hill-Rom, Inc.: Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.
- Recall number
- Z-2376-2016
- Recalling firm
- Hill-Rom, Inc.
- Product
- Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.
- Status
- Terminated
- Initiated
- 2016-06-14
- Posted
- 2016-08-09
- Terminated
- 2017-05-15
- Reason
- The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury.
- Root cause
- Device Design
- Action
- Hill-Rom sent an URGENT: MEDICAL DEVICE CORRECTION letter dated June 14, 2016 to the affected US and Canadian customers via email and courier, international customers. The letter included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The letter requests the number of beds in the consignees possession so that repair kits can be provided. If you have any questions regarding this safety notice, contact Hill-Rom Technical Support at 1-800-445-3720, your distributor, your Hill-Rom representative or contact us on www.hill-rom.com.
- Quantity
- 24262
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Canada, AUSTRALIA, CHINA, HONG KONG, THAILAND, POLAND, RUSSIAN FEDERATION, SLOVENIA, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, ICELAND, IRELAND, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, SPAIN, SWITZERLAND, and UNITED KINGDOM.
- Lot, serial or model codes
- All beds (serial numbers) in the 2080, 2082, S 280 and S 282 series are affected by this field action except for those without siderails. The models are as follows: 2080, 2082, 2080 C, 2080 E, 2080 FS, 2080 K, 2080 LP, 2080 MA, 2080 MB, 2080 S, 2080 SL, 2080 SONDER, 2082 SONDER, S280, S280 SONDER, S282, S282 SONDER.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 1824206
- Firm city
- Batesville
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as