The Binding Site Group, Ltd.: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Recall number
Z-2362-2026
Recalling firm
The Binding Site Group, Ltd.
Product
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
Status
Open, Classified
Initiated
2026-04-16
Posted
2026-06-09
Terminated
not on file
Reason
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
Root cause
Under Investigation by firm
Action
Firm notified customers via "Urgent Medical Device Correction" letter dated April 14, 2026. Customers were advised to confirm QC results are present in the QC results screen for the corresponding plate barcode after the completion of each plate measurement. If no QC result is displayed, do not release results from that plate and repeat the run ensuring QC material is properly loaded. The customer should also contact US Technical Support. When performing multi-plate runs while loading a previously used QC vials, ensure newly opened QCs are loaded onto Reagent Runner 2 as well, or discard and load newly opened QC vials into Reagent Runner 1. Retrospective review of previously released results is not considered necessary by the recalling firm. Please maintain awareness of the issue until Thermo Fisher Scientific identifies and implements a software enhancement that addresses the issue.
Quantity
1
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.
510(k) numbers
["K250159"]
PMA numbers
not on file
Product code
OTA
Firm FEI
3002808340
Firm city
Birmingham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register