The Binding Site Group, Ltd.: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
- Recall number
- Z-2362-2026
- Recalling firm
- The Binding Site Group, Ltd.
- Product
- Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
- Status
- Open, Classified
- Initiated
- 2026-04-16
- Posted
- 2026-06-09
- Terminated
- not on file
- Reason
- It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
- Root cause
- Under Investigation by firm
- Action
- Firm notified customers via "Urgent Medical Device Correction" letter dated April 14, 2026. Customers were advised to confirm QC results are present in the QC results screen for the corresponding plate barcode after the completion of each plate measurement. If no QC result is displayed, do not release results from that plate and repeat the run ensuring QC material is properly loaded. The customer should also contact US Technical Support. When performing multi-plate runs while loading a previously used QC vials, ensure newly opened QCs are loaded onto Reagent Runner 2 as well, or discard and load newly opened QC vials into Reagent Runner 1. Retrospective review of previously released results is not considered necessary by the recalling firm. Please maintain awareness of the issue until Thermo Fisher Scientific identifies and implements a software enhancement that addresses the issue.
- Quantity
- 1
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.
- 510(k) numbers
- ["K250159"]
- PMA numbers
- not on file
- Product code
- OTA
- Firm FEI
- 3002808340
- Firm city
- Birmingham
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28