Skip to the content

FDA Medical Device Recalls (openFDA)

Cadwell Industries Inc: Cascade IOMAX Cortical Module, REF: 190296-200

Recall number
Z-2349-2024
Recalling firm
Cadwell Industries Inc
Product
Cascade IOMAX Cortical Module, REF: 190296-200
Status
Open, Classified
Initiated
2024-05-17
Posted
2024-07-12
Terminated
not on file
Reason
Electroneurodiagnostic medical device may contain a defective printed circuit board which may cause damage to the other internal components resulting in unintended electrical stimulation.
Root cause
Process design
Action
On 5/17/2024, Cadwell Industries, Inc. issued "Urgent: Medical Device Recall" notification to affected consignees via E-Mail. Cadwell ask consignees to take the following actions: 1. Cease use of the device immediately and return all identified devices to Cadwell Industries, Inc. 2. Please complete the attached response form and return via email to quality@cadwell.com. 3. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred the IOMAX Cortical Module to. 4. Please complete acknowledgement and product replacement form.
Quantity
26 units
Distribution
US: IL CA OUS: Canada, Germany, India, Italy, Mexico, Poland, South Africa, Spain, Taiwan, United Kingdom
Lot, serial or model codes
UDI: 00840067100653/Serial Numbers:19029603AA0224001, 19029603AA0224002, 19029603AA0224003, 19029603AA0224004,19029603AA0224005, 19029603AA0224006, 19029603AA0224007, 19029603AA0224008, 19029603AA0224010, 19029603AA0224021, 19029603AA0224027, 19029603AA0324006, 19029603AA0324007, 19029603AA0324008, 19029603AA0324009, 19029603AA0324010, 19029603AA0424001, 19029603AA0424002, 19029603AA0424003, 19029603AA0424006, 19029603AA0424007, 19029603AA0424008, 19029603AA0424009, 19029603AA0424010, 19029603AA0524003, 19029603AA0524004
510(k) numbers
K162199
PMA numbers
not on file
Product code
GWF
Firm FEI
3020018
Firm city
Kennewick
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register