Howmedica Osteonics Corp.: Triathlon Prim CEM FXD BPLT 7, Catalog No. 5520B700

Recall number
Z-2329-2021
Recalling firm
Howmedica Osteonics Corp.
Product
Triathlon Prim CEM FXD BPLT 7, Catalog No. 5520B700
Status
Terminated
Initiated
2021-07-01
Posted
not on file
Terminated
2023-05-17
Reason
The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.
Root cause
Under Investigation by firm
Action
On July 1, 2021, the firm notified customers via Urgent Medical Device Recall. The letter listed the hazards and harms potentially associated with the recalled product and also listed mitigation factors. Patients treated with an affected product should continue to be followed per the normal protocol established by his or her surgeon(s). There are no recommended changes to the frequency of the standard follow-up care protocol. It should be noted that the failure of the insert to fully seat, may not be evident on x-ray unless the insert has dislodged. Actions Needed: Our records indicate that you may have received the affected product(s). It is Strykers responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important follow up communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list. Return the devices back to Stryker. Customers should complete and sign the included customer response form and return it to the recalling firm. Hospitals should contact their local sales office or Stryker sales representative directly for product returns and inventory questions. Customers who have further distributed the product should forward should notify downstream accounts and return all affected devices to the recalling firm. For questions pertaining to the recall, please contact Customer Service at (201) 831-5000, or email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
Quantity
14 US; 2 OUS
Distribution
Worldwide distribution - US Nationwide distribution in the states AR, CO, MN, MS, NJ, OK, PA, TX and the countries of China and Italy.
Lot, serial or model codes
Lot JHS2H
510(k) numbers
["K141056"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register