Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management

Recall number
Z-2317-2019
Recalling firm
Teleflex Medical
Product
Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
Status
Terminated
Initiated
2019-07-30
Posted
2019-09-09
Terminated
2022-02-15
Reason
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
Root cause
Process control
Action
Teleflex sent an Urgent Medical Device Recall Notification letter dated July 30, 2019 to affected customers. The letter explains the problem and requests the consignee discontinue use and quarantine of any of the recalled products then to notify Teleflex. The firm is requesting the return of the affected products. For questions, contact your local sales representative or Customer Service at 1-866-396-2111.
Quantity
77136 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
Batch numbers:  73K1600156 73K1600735 73L1600501 73A1700120 73B1700321 73C1700425 73D1700630 73E1700504 73G1700146 73H1700053 73J1700307 73K1700085 73B1800223 73B1800371 73C1800372 73D1800655 73E1800579 73G1800627 73H1800203 73H1800429 73H1800758 73J1800118 73J1800696
510(k) numbers
K822082
PMA numbers
not on file
Product code
BTR
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register