Skip to the content

FDA Medical Device Recalls (openFDA)

Grace Medical, Inc.: TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis

Recall number
Z-2308-2026
Recalling firm
Grace Medical, Inc.
Product
TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis
Status
Open, Classified
Initiated
2026-04-03
Posted
2026-06-04
Terminated
not on file
Reason
Due to incorrect functional length on device labeling.
Root cause
Process control
Action
On 04/03/2026, the firm emailed a "RECALL NOTICE" to it single customer informing them that the firm determined the pistons were manufactured with an incorrect functional length. Devices are labeled as 3.75 mm but have an actual functional length of 4.00 mm. The customer was instructed to: " Immediately discontinue use and quarantine affected product " Return affected product using the provided RMA " If any devices have been used, provide the date(s) of surgery on the response form " Complete and return the enclosed Recall Response Form For questions - contact Grace Medical at 901-386-0990 or Thomas.fearnley@gracemedical.com.
Quantity
7
Distribution
U.S Nationwide distribution in the state of MD.
Lot, serial or model codes
Model/Catalog Number: 409-375 UDI code: (01)00844505000772(10)122927(17)310301 Lot Number: 122927
510(k) numbers
["K060518"]
PMA numbers
not on file
Product code
ETB
Firm FEI
1000022662
Firm city
Bartlett
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06