Skip to the content

FDA Medical Device Recalls (openFDA)

Bayer Medical Care, Inc.: Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;

Recall number
Z-2296-2026
Recalling firm
Bayer Medical Care, Inc.
Product
Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
Status
Open, Classified
Initiated
2026-04-24
Posted
2026-06-03
Terminated
not on file
Reason
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
Root cause
Process control
Action
On April 24, 2026, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Customers were instructed to immediately halt the use of the affected product and review inventory for impacted product and quarantine. A Return Goods Authorization will be sent to return any remaining affected inventory. Place the affected product in a box and label the outside of the container with the RGA number provided and RECALL ID: SA-2026-RAD-03 in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return the affected product as soon as possible. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Quantity
11,300 units (1650 US, 9650 OUS)
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
Lot, serial or model codes
Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648; Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802;
510(k) numbers
["K050456"]
PMA numbers
not on file
Product code
DXT
Firm FEI
2520313
Firm city
Indianola
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register