Eizo Corporation: RadiForce GX560 monochrome LCD monitor, models 0FTD2347 and 0FTD2379 - Product Usage: is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

Recall number
Z-2293-2020
Recalling firm
Eizo Corporation
Product
RadiForce GX560 monochrome LCD monitor, models 0FTD2347 and 0FTD2379 - Product Usage: is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Status
Open, Classified
Initiated
2019-07-30
Posted
not on file
Terminated
not on file
Reason
Under certain conditions, a marble pattern infrequently appears on the monitor.
Root cause
Device Design
Action
EIZO Japan announced the recall on their website on 7/30/2019. EIZO U.S. began contacting their consignees via email and phone (one as early as 8/7/2019, some on 10/18/2019, but the majority on 11/26/2019) and began updating consignee monitors with the firmware. The notification requested the consignee contact EIZO U.S. for assistance with the firmware update.
Quantity
96 monitors dist. to the U.S.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Louisiana, Maine, Massachusetts, Michigan, Missouri, Montana, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Dakota, Tennessee, Texas, Virginia, Washington, and Wisconsin. Distribution was also made to Puerto Rico. There was no government/military consignees. The countries of Canada, Australia, Austria, Belgium, Brazil, China, Czech Republic, Egypt, France, Germany Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Lebanon, Netherlands, Pakistan, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
RadiForce GX560 - 1.0003 or earlier
510(k) numbers
["K181609"]
PMA numbers
not on file
Product code
PGY
Firm FEI
3000237660
Firm city
Hakusan
Firm state
not on file
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register