Skip to the content

FDA Medical Device Recalls (openFDA)

Brainlab AG: BrainLab Radiotherapy Treatment Planning Software; Catalog number 20610 - Radiosurgery 3.0 Catalog number 20620 - Radiosurgery 3.5 and Catalog number 20630 - Circular ARC SRS/SRT Planning. The software is intended for use in stereotactic, conformal, computer planned, LINAC based radiation treatment of cranial, head and neck, and extracranial lesions. It is intended to be used by experienced and trained health professionals.

Recall number
Z-2251-2009
Recalling firm
Brainlab AG
Product
BrainLab Radiotherapy Treatment Planning Software; Catalog number 20610 - Radiosurgery 3.0 Catalog number 20620 - Radiosurgery 3.5 and Catalog number 20630 - Circular ARC SRS/SRT Planning. The software is intended for use in stereotactic, conformal, computer planned, LINAC based radiation treatment of cranial, head and neck, and extracranial lesions. It is intended to be used by experienced and trained health professionals.
Status
Terminated
Initiated
2009-08-12
Posted
2009-09-20
Terminated
2011-03-16
Reason
Failure to conduct important safety checks when using the BrainLab radiotherapy treatment planning software in combination with BrainLab conical collimators could result in unintended radiation outside the conical shaped field, which may lead to serious injury of the patient.
Root cause
Other
Action
BrainLAB AG issued an "Urgent Field Safety Notice" letter dated August 12, 2009 informing all customers of the BrainLab radiotherapy planning software using conical collimators of the important safety checks to avoid the failure to apply and verify an adequate jaw-shaped field size and field position at the LINAC for the used conical collimator(s) or the failure to mount and verify the correct, intended conical collimator. Consignees were instructed on proper steps to take to avoid further issues. For further questions, contact BrainLAB AG by phone at 1-800-597-5911. *** On February 1, 2010, BrainLab sent supplemental letters to 50 of the accounts who received the August 12, 2009 letter and have Novalis Dose Delivery Systems (excluding all Novalis Tx systems). The accounts were informed that in addition to the safety checks listed in the August letter, BrainLab will be installing a mandatory software modification on their systems that limits the field size to a maximum of 4x4 cm2 when the conical collimators are inserted. The accounts would be contacted by a service representative within the next six weeks to schedule the update. Any questions were directed to BrainLab at 1-800-597-5911. ***
Quantity
230 units
Distribution
Nationwide.
Lot, serial or model codes
All versions of BrainSCAN or iPlan RT Dose software used with BrainLAB conical collimators.
510(k) numbers
K002509, K080888, K920879, K994413
PMA numbers
not on file
Product code
MUJ
Firm FEI
3002619595
Firm city
Kirchheim B. Muenchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register