Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.

Recall number
Z-2244-2012
Recalling firm
Medtronic Navigation, Inc.
Product
Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.
Status
Terminated
Initiated
2012-07-11
Posted
2012-08-22
Terminated
2016-06-06
Reason
A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.
Root cause
Component change control
Action
Medtronic Navigation issued an "Urgent Field Safety Notice" to consignees dated 7/11/12. The problem was described and recommended actions were provided. For each affected O-arm Imaging System, a Medtronic service representative will visit the site and replace the 500VDC relay on the system and test the O-arm Imaging System.
Quantity
17 units
Distribution
Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain.
Lot, serial or model codes
Serial Numbers: 109R, 118R, 124, 136, 159 ,170, 218, 222, 223, 236, 239 , 299, 309R, 402, 410, 412, 414.
510(k) numbers
["K092564"]
PMA numbers
not on file
Product code
OXO
Firm FEI
3004785967
Firm city
Littleton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register